Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/107019
Title: Simple, Immediate and Calibration-Free Cyclotron Proton Beam Energy Determination Using Commercial Targets
Authors: Carmo, Sérgio J. C. do 
Oliveira, Pedro M. 
Alves, Francisco 
Keywords: cyclotron; stack-foil; monitor reaction
Issue Date: 2019
Publisher: MDPI
Serial title, monograph or event: Instruments
Volume: 3
Issue: 1
Abstract: This work presents a simple method for determining the energy of the proton beam in biomedical cyclotrons, using no additional experimental set-up and only materials from radioisotope routine productions that are therefore available on-site. The developed method requires neither absolute efficiency calibration nor beam current measurements, thus avoiding two major sources of uncertainty. Two stacks composed of natural titanium thin foils, separated by an energy degrader of niobium, were mounted in a commercial target and irradiated. The resulting activities of 48V were assessed by a HPGe spectrometer.
URI: https://hdl.handle.net/10316/107019
ISSN: 2410-390X
DOI: 10.3390/instruments3010020
Rights: openAccess
Appears in Collections:I&D ICNAS - Artigos em Revistas Internacionais