Utilize este identificador para referenciar este registo: https://hdl.handle.net/10316/106100
Título: Multi-Material Implants for Temporomandibular Joint Disc Repair: Tailored Additive Manufacturing Production
Autor: Moura, Carla 
Trindade, Daniela
Vieira, Milena 
Francisco, Luís
Ângelo, David Faustino
Alves, Nuno
Palavras-chave: temporomandibular joint disc; processing conditions; Poly(+-caprolactone); Poly(ethylene glycol) diacrylate; multi-material structures
Data: 2020
Editora: Frontiers Media S.A.
Projeto: UID/Multi/04044/2019 
PAMI – ROTEIRO/0328/2013 (No. 022158) 
MATIS (CENTRO-01-0145- FEDER-000014 – 3362) 
BIODISCUS (CENTRO-01-0247- FEDER-039969) 
Título da revista, periódico, livro ou evento: Frontiers in Bioengineering and Biotechnology
Volume: 8
Resumo: Temporomandibular disorders (TMD) affect a substantial percentage of the population, and the resources spent on their treatment are considerable. Despite the worldwide efforts around Tissue Engineering of the temporomandibular joint (TMJ) disc, a proper implant offering a long-term solution for TMD was not yet developed. To contribute to these efforts, this work is focused on the research and development of implants for TMJ disc regeneration. Scaffolds and hydrogels mimicking the TMJ disc of black Merino sheep were produced using different materials, poly(ε-caprolactone) (PCL) and poly(ethylene glycol) diacrylate (PEGDA), and as a multi-material structure. Different parameters of the scaffold manufacturing were assessed: the influence of processing temperatures, filament diameter, and biological environment. Moreover, two multi-material approaches were also assessed, scaffold with a hydrogel shell and scaffold with a hydrogel core. It was found that increasing temperature, the scaffolds' porosity decreases, increasing their compressive modulus. Decreasing the filament size (300 to 200 μm) decreases the compressive modulus to almost half of the initial value. Scaffolds with 200 μm filaments are the ones with a closer modulus to the native disc and their properties are maintained under hydrated conditions. The introduction of a hydrogel core in these scaffolds presented better mechanical properties to TMJ disc substitution.
URI: https://hdl.handle.net/10316/106100
ISSN: 2296-4185
DOI: 10.3389/fbioe.2020.00342
Direitos: openAccess
Aparece nas coleções:I&D CEMMPRE - Artigos em Revistas Internacionais

Mostrar registo em formato completo

Citações SCOPUSTM   

15
Visto em 15/jul/2024

Citações WEB OF SCIENCETM

12
Visto em 2/jul/2024

Visualizações de página

61
Visto em 16/jul/2024

Downloads

41
Visto em 16/jul/2024

Google ScholarTM

Verificar

Altmetric

Altmetric


Este registo está protegido por Licença Creative Commons Creative Commons