Título: Operation of gas proportional scintillation counters in a low charge multiplication regime
Autor: Borges, F. I. G. M. 
Santos, J. M. F. dos 
Dias, T. H. V. T. 
Santos, F. P. 
Rachinhas, P. J. B. M. 
Conde, C. A. N. 
Data: 1999
Citação: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 422:1-3 (1999) 321-325
Resumo: The operation of a xenon filled gas proportional scintillation counter under low charge multiplication gains is discussed. It is shown that the best energy resolution for X-rays is obtained for a reduced electric field, E/p, of about 8 V cm-1 Torr-1, leading to a charge gain (for inter-grid distance of 5 mm) of about 1.05, as calculated by a Monte Carlo simulation method. A discussion is presented concerning the dependence of the E/p values for better detector performance, on the inter-grid distance.