Utilize este identificador para referenciar este registo:
https://hdl.handle.net/10316/4519
Título: | Gas proportional scintillation counters with a CsI-covered microstrip plate UV photosensor for high-resolution X-ray spectrometry | Autor: | Veloso, J. F. C. A. Santos, J. M. F. dos Conde, C. A. N. |
Palavras-chave: | Gas proportional scintillation counter; CSI photocathode | Data: | 2001 | Citação: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 457:1-2 (2001) 253-261 | Resumo: | A xenon-filled gas proportional scintillation counter for X-ray spectrometry is described. The detector uses a microstrip plate covered with a thin CsI film as a photosensor rather than the usual photomultiplier tube. The achieved energy resolution for 5.9 keV X-rays is 12% when the microstrip is operated within the xenon volume and 10.5% when the microstrip is operated in a P-10 environment, separated from the xenon volume by a quartz window. | URI: | https://hdl.handle.net/10316/4519 | DOI: | 10.1016/S0168-9002(00)00746-4 | Direitos: | openAccess |
Aparece nas coleções: | FCTUC Física - Artigos em Revistas Internacionais |
Ficheiros deste registo:
Ficheiro | Descrição | Tamanho | Formato | |
---|---|---|---|---|
file3484bb67eef3497b9720eacf6880d567.pdf | 463.04 kB | Adobe PDF | Ver/Abrir |
Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.