Title: Energy linearity response of CZT detectors to X-rays in the region of the Zn, Cd and Te K-absorption edges: experimental results
Authors: Freitas, E. D. C. 
Cardoso, João M. 
Fernandes, L. M. P. 
Lopes, J. A. M. 
Veloso, J. F. C. A. 
Santos, J. M. F. dos 
Keywords: Energy linearity; K-edges; X-ray detectors
Issue Date: 2003
Keywords: Energy linearity; K-edges; X-ray detectors
Issue Date: 2003
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 515:3 (2003) 618-623
Abstract: The response of a cadmium zinc telluride (CZT) detector to X-rays with energies around the Zn, Cd and Te K-absorption edges was investigated. The energy non-linearity in the detector response at the K-edges is less than 0.1%, within the experimental errors. No abrupt variation in both the w-value and the detector energy resolution was observed at the K-edges, a behaviour consistent with the absence of intrinsic non-linearity effects in CZT. Within the accuracy of our measurements, we conclude that there are no non-linearity effects in CZT semiconductors at the Zn, Cd and Te K-edges.
URI: http://hdl.handle.net/10316/4461
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
filec6897aec8ee243f5a529d739e287b67d.pdf128.32 kBAdobe PDFView/Open
Show full item record

Page view(s)

74
checked on Aug 14, 2019

Download(s)

41
checked on Aug 14, 2019

Google ScholarTM

Check


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.