Utilize este identificador para referenciar este registo: https://hdl.handle.net/10316/4399
Título: Spatial resolution on a small animal RPC-PET prototype operating under magnetic field
Autor: Blanco, A. 
Carolino, N. 
Correia, C. M. B. A. 
Fazendeiro, L. 
Ferreira, Nuno C. 
Marques, M. F. Ferreira 
Marques, R. Ferreira 
Fonte, P. 
Gil, C. 
Macedo, M. P. 
Palavras-chave: Gaseous detectors; magnetic field
Data: 2006
Citação: Nuclear Physics B - Proceedings Supplements. 158:(2006) 157-160
Resumo: It has been demonstrated in previous work that the RPC-PET technology is able to deliver radioisotope image resolutions approaching the physical limits of the PET principle. Here we study, by simulation, the effect of the magnetic field on the positron range to evaluate whether the spatial resolution of the RPC-PET could be improved by using an intense magnetic field. Six positron emitters of interest to small animal PET imaging (18F, 11C, 15O, 68G, 62Cu and 86Y) are considered. Results suggest that a three-fold improvement on the spatial resolution may be obtained under a magnetic field of 10 T for the higher energy radioisotopes like 86Y or 62Cu, and by about 20% for the lower energy ones, like 18F or 11C.
URI: https://hdl.handle.net/10316/4399
DOI: 10.1016/j.nuclphysbps.2006.07.006
Direitos: openAccess
Aparece nas coleções:FCTUC Física - Artigos em Revistas Internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato
filec9743bb8c8ba4d88bf4084a03350abaa.pdf233.27 kBAdobe PDFVer/Abrir
Mostrar registo em formato completo

Citações SCOPUSTM   

3
Visto em 22/abr/2024

Citações WEB OF SCIENCETM

3
Visto em 2/abr/2024

Visualizações de página 50

407
Visto em 23/abr/2024

Downloads 50

357
Visto em 23/abr/2024

Google ScholarTM

Verificar

Altmetric

Altmetric


Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.