Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4361
Title: High-rate operation of the Micro-Hole and Strip Plate gas detector
Authors: Veloso, J. F. C. A. 
Caldas, C. C. 
Oliveira, C. A. B. 
Azevedo, C. D. R. 
Santos, J. M. F. dos 
Breskin, A. 
Chechik, R. 
Keywords: Count rate
Issue Date: 2007
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 580:1 (2007) 362-365
Abstract: We report the performance of the Micro-Hole-and-Strip-Plate (MHSP) operated in xenon, under high-flux irradiation of a 30-kV X-ray Cu-tube. Stable operation in pulse-counting mode for gas gains above 104 was measured up to counting rates of 5×105 Hz/mm2. A gain variation of less than 5% was observed for gains of 104, while the energy resolution for 8-keV X-rays degrades from ~16% to ~18%. No visible degradation was observed on the MHSP irradiated area.
URI: https://hdl.handle.net/10316/4361
DOI: 10.1016/j.nima.2007.05.179
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais