Please use this identifier to cite or link to this item: http://hdl.handle.net/10316/4503
Title: Two-dimensional readout in a liquid xenon ionisation chamber
Authors: Solovov, V. 
Chepel, V. 
Pereira, A. 
Lopes, M. I. 
Marques, R. Ferreira 
Policarpo, A. J. P. L. 
Keywords: Liquid xenon; Ionisation chamber; Gamma-ray detection
Issue Date: 2002
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 477:1-3 (2002) 184-190
Abstract: A two-dimensional readout with metal strips deposited on both sides of a glass plate is investigated aiming to assess the possibility of its use in a liquid xenon ionisation chamber for positron emission tomography. Here, we present results obtained with an [alpha]-source. It is shown that position resolution of [less-than-or-equals, slant]1 mm, fwhm, can be achieved for free charge depositions equivalent to those due to [gamma]-rays with energy from 220 down to 110 keV.