Please use this identifier to cite or link to this item:
https://hdl.handle.net/10316/4472
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Freitas, D. S. A. P. | - |
dc.contributor.author | Veloso, J. F. C. A. | - |
dc.contributor.author | Monteiro, C. M. B. | - |
dc.contributor.author | Santos, J. M. F. dos | - |
dc.contributor.author | Conde, C. A. N. | - |
dc.date.accessioned | 2008-09-01T11:07:41Z | - |
dc.date.available | 2008-09-01T11:07:41Z | - |
dc.date.issued | 2003 | en_US |
dc.identifier.citation | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 505:1-2 (2003) 228-232 | en_US |
dc.identifier.uri | https://hdl.handle.net/10316/4472 | - |
dc.description.abstract | The gas proportional scintillation counter/microstrip gas chamber (GPSC/MSGC) hybrid detector for X-ray spectrometry is described. The detector uses a CsI-coated microstrip plate placed in direct contact with the gas-filling as the photosensor readout for the GPSC scintillation substituting for the photomultiplier tube (PMT). Usable photosensor maximum gain is limited by optical positive feedback due to the additional scintillation produced in the electron avalanche process at the MSP anodes, in the absence of quenching. A low-photoelectron collection efficiency is achieved in the gas atmosphere, resulting in a scintillation conversion efficiency that is about a factor of 5 lower than that achieved with PMT-based GPSCs. However, energy resolutions of 11% for 5.9 keV X-rays are achieved with this detector. | en_US |
dc.description.uri | http://www.sciencedirect.com/science/article/B6TJM-486BN6Y-M/1/4da741da1d82059bc939bc52156b6569 | en_US |
dc.format.mimetype | aplication/PDF | en |
dc.language.iso | eng | eng |
dc.rights | openAccess | eng |
dc.subject | Gas scintillation counter | en_US |
dc.subject | CsI photocathode | en_US |
dc.title | The gas proportional scintillation counter/microstrip gas chamber hybrid detector | en_US |
dc.type | article | en_US |
dc.identifier.doi | 10.1016/S0168-9002(03)01058-1 | - |
item.fulltext | Com Texto completo | - |
item.openairecristype | http://purl.org/coar/resource_type/c_18cf | - |
item.grantfulltext | open | - |
item.languageiso639-1 | en | - |
item.openairetype | article | - |
item.cerifentitytype | Publications | - |
crisitem.author.researchunit | LIBPhys - Laboratory for Instrumentation, Biomedical Engineering and Radiation Physics | - |
crisitem.author.researchunit | LIBPhys - Laboratory for Instrumentation, Biomedical Engineering and Radiation Physics | - |
crisitem.author.parentresearchunit | University of Coimbra | - |
crisitem.author.parentresearchunit | University of Coimbra | - |
crisitem.author.orcid | 0000-0002-7107-7203 | - |
crisitem.author.orcid | 0000-0002-1912-2804 | - |
crisitem.author.orcid | 0000-0002-8841-6523 | - |