Título: Quality control of GEM detectors using scintillation techniques
Autor: Fraga, F. A. F. 
Fetal, S. T. G. 
Marques, R. Ferreira 
Policarpo, A. J. P. L 
Palavras-chave: GEM detectors; Microstructures
Data: 2000
Citação: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 442:1-3 (2000) 417-422
Resumo: Non-destructive quality control of microstructures at the manufacturing stage is an important issue in the foreseen use of huge numbers of such gaseous detectors in the future high luminosity colliders. In this work we report on the use of the scintillation light emitted by the avalanches in GEM channels for checking defects in the foils. The test system is described and data on the relative efficiency of several gaseous mixtures are presented. The foil images obtained with a low-noise CCD system are analysed and compared with the optical images obtained with an industrial inspection system of high magnification. The validity of this test method is established and possible extensions of its use are discussed.
URI: https://hdl.handle.net/10316/4535
DOI: 10.1016/S0168-9002(99)01266-8
Direitos: openAccess
Aparece nas coleções:FCTUC Física - Artigos em Revistas Internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato
filed6aba242340f4f6a836f298eb810d12a.pdf490.68 kBAdobe PDFVer/Abrir
Mostrar registo em formato completo

Citações SCOPUSTM   

16
Visto em 15/abr/2024

Citações WEB OF SCIENCETM

11
Visto em 2/abr/2024

Visualizações de página 50

409
Visto em 23/abr/2024

Downloads 50

618
Visto em 23/abr/2024

Google ScholarTM

Verificar

Altmetric

Altmetric


Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.