Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4518
Title: Enhancing the ratio of fluorescence to bremsstrahlung radiation in X-ray tube spectra
Authors: Távora, L. M. N. 
Morton, E. J. 
Gilboy, W. B. 
Keywords: Fluoresence radiation; Bremsstrahlung; X-ray tube spectra
Issue Date: 2001
Citation: Applied Radiation and Isotopes. 54:1 (2001) 59-72
Abstract: This paper describes techniques that can be used to improve the ratio of fluorescence to bremsstrahlung radiation (F/B) in X-ray tube spectra. Firstly, an extension of the EGS4 code system is used to evaluate the impact of the substrate in thin target applications, in terms of the yield of bremsstrahlung photons produced. The choice of materials to filter X-ray tube spectra, and its effect in the F/B and the tube efficiency is discussed. The characteristics of spectra produced in transmission tubes with different target thicknesses, substrates and tube voltages are also presented.
URI: https://hdl.handle.net/10316/4518
DOI: 10.1016/S0969-8043(00)00165-2
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
file62f2a73a14574371b85865fca0ef98e6.pdf522.49 kBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

2
checked on Apr 22, 2024

WEB OF SCIENCETM
Citations

3
checked on Apr 2, 2024

Page view(s)

231
checked on Apr 23, 2024

Download(s)

280
checked on Apr 23, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.