Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4458
Title: X-ray spectrometry with Peltier-cooled large area avalanche photodiodes
Authors: Fernandes, L. M. P. 
Lopes, J. A. M. 
Santos, J. M. F. dos 
Conde, C. A. N. 
Keywords: Avalanche photodiode; X-ray spectrometry; Peltier-cooling
Issue Date: 2004
Citation: Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 213:(2004) 267-271
Abstract: Performance characteristics of the response of a Peltier-cooled large-area avalanche photodiode are investigated. Detector gain, energy linearity, energy resolution and minimum detectable energy are studied at different operation temperatures. Detector energy resolution and lowest detectable X-ray energy present a strong improvement as the operation temperature is reduced from 25 to 15 °C and slower improvements are achieved for temperatures below 10 °C.
URI: https://hdl.handle.net/10316/4458
DOI: 10.1016/S0168-583X(03)01607-0
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
fileb92d4329efab44aaa26af62c12e64a51.pdf308.73 kBAdobe PDFView/Open