Título: Ceramic high-rate timing RPCs
Autor: Lopes, L. 
Marques, R. Ferreira 
Fonte, P. 
Hennetier, L. 
Pereira, A. 
Correia, A. M. Sousa 
Palavras-chave: High-rate; Ceramic electrode
Data: 2006
Citação: Nuclear Physics B - Proceedings Supplements. 158:(2006) 66-70
Resumo: Following some previous work, we report here considerable improvements on the counting rate capability of timing RPCs by the use of ceramic electrodes with a resistivity of 109 [Omega][dot operator]cm. The X-ray sensitivity of the detector depends linearly on the counting rate with a slope of 9% per 100 kHz/cm2, free from charge depletion effects, while keeping a timing accuracy, measured with 511 keV synchronous photon pairs, around 90 ps [sigma] up to 75 kHz/cm2.
URI: https://hdl.handle.net/10316/4397
DOI: 10.1016/j.nuclphysbps.2006.07.017
Direitos: openAccess
Aparece nas coleções:FCTUC Física - Artigos em Revistas Internacionais